PULSE
S3023P
Single channel High-Voltage Semiconductor Pulse Generator Unit S3023P
Semight InstrumentsS3023P It is a single channel High-Voltage Semiconductor Pulse Generator Unit that can output adjustable high-precision high-voltage pulse sources.
Features
High voltage range
±40V,Meet NVM testing technical conditionsHigh voltage with fast rate of change
1000V/us,A high-voltage pulse source that can generate faster edgesStandard pulse mode
three-level pulse, period:40ns~10sDelay adjustable
time:0~9.999s,resolution ratio:2.5nsHigh-speed measurement
500MS/s,Timely diagnosis of load impedanceAdjustable pulse count
range:1~100,000个Minimum sampling period for voltage detection
5usLevel timing trigger
TTLLevel, pulse cycle synchronizationFeatures and Benefits
Working conditions:
Temperature 23 ℃± 5 ℃
Project | Test conditions | Technical indicators | Remarks |
Voltage amplitude |
50 Ω load | -20V~+20V |
|
Open circuit | -40V~+40V | ||
Voltage amplitude resolution |
50 Ω load |
0.2mV | |Vout | ≤ 5V |
0.8mV | 0.5V ≤ | Vout | ≤ 20V | ||
|
0.4mV | | Vout | ≤ 10V | |
1.6mV | 10V ≤ | Vout | ≤ 40V | ||
Output interface |
|
SMA |
|
Internal resistance |
|
50 Ω ± 1% |
|
Output short-circuit current |
|
800mA peak (400mA mean) |
|
Output overshoot (ringing) | 50 Ω load | ± (5%+20mV) |
|
Project | Sub projects | Technical indicators | Remarks |
Frequency range |
|
0.1~33MHz |
|
Pulse period |
Programmable range | 40ns~10s |
|
Resolution | 10ns |
|
|
Minimum cycle | 100ns | Vamp ≤ 10V | |
Accuracy | ± 1% (0.01%+200ps) |
|
|
Pulse width |
Programmable range | 20ns~(pulse period -10ns) |
|
Resolution |
2.5ns | Vamp< 10V | |
20ns | Vamp>10V | ||
Minimum value (10%~90% edge) | 120ns |
|
|
Accuracy | ± (3%+2ns) |
|
|
Pulse process time (Tr and Tf) |
Programmable range | 20ns~400ms |
|
Resolution |
2ns | Tr+Tf ≤ 10us | |
20ns | Tr+Tf>10us | ||
Minimum value (10%~90% edge) |
30ns | Vamp ≤ 5V | |
35ns | Vamp ≤ 10V | ||
40ns | Vamp ≤ 20V | ||
60ns | Vamp>20V | ||
Accuracy |
-5%~+5%+25ns | Vamp ≤ 10V | |
-5%~+5%+40ns | Vamp ≤ 20V |
Advanced features:
Project | Technical indicators | Test conditions |
Pulse width jitter | 0.001%+150ps |
|
Pulse period jitter | 0.001%+150ps |
|
Maximum voltage swing rate | 1000V/us | Vamp=40V (50 Ω Load) |
Output noise | 20mVrms |
|
Voltage monitoring | SPGU monitoring and testing unit terminal voltage |
|
Measurement accuracy | ± (0.1%+25mV) |
|
Measurement resolution | 50uV |
|