S3012H
Single-Channel Precision SMU
The S3012H precision source meter is compact and cost-effective bench-top Source/Measure Units (SMUs) with the capability to source and measure both voltage and current. These capabilities make the S3012H ideal for a wide variety of IV (current versus voltage) measurement tasks that require both high resolution and accuracy.
Features
High range
Range: ±200 V, ±3 A (DC), ±10 A (pulse)High resolution
Minimum measurement resolution up to 100 fA/100 nVHigh sample rate
Up to 1M ADC sampling rate can be supportedhardware synchronization
Single-machine multi-machine synchronization, pure hardware high-speed synchronization, can realize multi-channel low-latency synchronizationFeatures and Benefits
5 functions in one
Power source1. The first and third quadrants are the source: the actual polarity of the output V/I follows the source setting;
2. The second and fourth quadrants are the load: CC and CV cooperate, when the load is used, the load setting polarity is opposite to the source polarity;
Various devices can be tested
Capture more measurement data
♦ 6½-digit digital resolution: The accuracy is equivalent to a 6½-digit digital multimeter;Rich scanning functions
Voltage source specifications
|
Range | Programming resolution |
Accuracy (1 Year) ± (% reading+ offset) |
Typical Noise (RMS) 0.1 Hz-10Hz |
|||
±200 V | 1 mV | 0.02% + 30 mV | 1.5 mV | ||||
±20 V | 100 μV | 0.02% + 2 mV | 160 μV | ||||
±6 V | 50 μV | 0.02% + 500 μV | 36 μV | ||||
±200 mV | 1 μV | 0.02% + 120 μV | 4 μV | ||||
Temperature coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) | ||||||
Maximum output power | 30W: ±20V@1.5A; ±200V@0.1A; 18W: ±6 V@3A | ||||||
Settling time | <800 μs (typical) | ||||||
Overshoot | <±0.1% (Typical.Normal.Step is 10 % to 90 % range, full range,resistive load) | ||||||
Noise 10Hz-20MHz | 6 V voltage source, 3 A resistive load, <3 mVrms |
Current source specifications
|
Range | Programming resolution |
Accuracy (1 Year) ± (% reading+ offset) |
Typical Noise (RMS) 0.1 Hz-10 Hz |
|||
±10 A1 | 50 μA | 0.4% + 40 mA | NA | ||||
±3 A | 15 μA | 0.05% + 2 mA | 40 μA | ||||
±1.5 A | 10 μA | 0.02% + 500 μA | 20 μA | ||||
±150 mA | 1 μA | 0.02% + 25 μA | 5 μA | ||||
±15 mA | 100 nA | 0.02% + 6 μA | 700 nA | ||||
±1.5 mA | 10 nA | 0.02% + 250 nA | 16 nA | ||||
±150 μA | 1 nA | 0.02% + 25 nA | 1 nA | ||||
±15 μA | 100 pA | 0.02% + 3 nA | 140 pA | ||||
±1.5 μA | 10 pA | 0.03% + 450 pA | 25 pA | ||||
±150 nA | 1 pA | 0.05% + 250 pA | 5 pA | ||||
Temperature coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) | ||||||
Maximum output power | 30W: ±20V@1.5A, ±200V@0.1A; 18W: ±6 V@3A | ||||||
Settling time | <500 μs (typical) | ||||||
Overshoot | <±0.1% (Typical.Normal.Step is 10 % to 90 % range, full range, resistive load) |
1、10 A range is available only for pulse mode, accuracy specifications for 10 A range are typical.
Voltage measurement specifications
|
Range | Measurement resolution | Accuracy (1 Year) ± (% reading+ offset) | ||||
±200 V | 100 μV | 0.02% + 30 mV | |||||
±20 V | 10 μV | 0.02% + 2 mV | |||||
±6 V | 1 μV | 0.02% + 500 μV | |||||
±200 mV | 100 nV | 0.02% + 120 μV | |||||
Temperature coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) |
Current measurement specifications
|
Range | Measurement resolution | Accuracy (1 Year)± (% reading+ offset) | ||||
±10 A1 | 10 μA | 0.4% + 25 mA | |||||
±3 A | 10 μA | 0.05% + 2 mA | |||||
±1.5 A | 1 μA | 0.02% + 500 μA | |||||
±150 mA | 100 nA | 0.02% + 25 μA | |||||
±15 mA | 10 nA | 0.02% + 6 μA | |||||
±1.5 mA | 1 nA | 0.02% + 250 nA | |||||
±150 μA | 100 pA | 0.02% + 25 nA | |||||
±15 μA | 10 pA | 0.02% +3 nA | |||||
±1.5 μA2 | 1 pA | 0.03% + 450 pA | |||||
±150 nA2 | 100 fA | 0.05% + 250 pA | |||||
Temperature coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) |
Resistance measurement specifications (4W)
|
Range | Measurement resolution | Test current | Typical accuracy(1 Year)±(% reading+ offset) | |||
1 Ω | 1 μΩ | 1.5 A | 0.073% +0.3334 mΩ | ||||
10 Ω | 10 μΩ | 150 mA | 0.057% + 3.334 mΩ | ||||
100 Ω | 100 μΩ | 15 mA | 0.08% + 33.34 mΩ | ||||
1 kΩ | 1 mΩ | 1.5 mA | 0.057% + 333.4 mΩ | ||||
10 KΩ | 10 mΩ | 150 μA | 0.057% + 3.334 Ω | ||||
100 kΩ | 100 mΩ | 15 μA | 0.06% + 33.34 Ω | ||||
1 MΩ | 1 Ω | 1.5 μA | 0.06% + 333.4 Ω | ||||
10 MΩ | 10 Ω | 0.15 μA | 0.35% + 3.334 kΩ | ||||
100 MΩ | 100 Ω | 0.05 μA | 0.95% + 10 kΩ | ||||
Temperature coefficient
Source I mode, manual Ohm measurement
(4-wire)
|
±(0.15 × accuracy)/°C
(0℃-18℃,28℃-50℃) Total error = Vmeas/Isrc = R reading x (gain error % of V range + gain error % of I range + offset error of I source range/Isrc value %) + (offset error of V measure range/Isrc value) Example:I source value=1.5A at 1.5A range V measure range=6V range Total error (% reading + offset) = (0.02%+0.02%+500μA/1.5A+(500μV/1.5A) ≈ 0.073%+0.3334mΩ |
Pulse source specifications (4W)
Minimum programmable pulse width | 100 μs | ||||||
Pulse width programming resolution | 1 μs | ||||||
Pulse width programming accuracy | ±10 μs | ||||||
Pulse width jitter | 2 μs | ||||||
Pulse width definition | The time from 10 % leading to 90 % trailing edge as follows |
Item | Maximums | Maximum pulse width | Maximum duty cycle | ||||
1 | 0.15 A/200 V | DC,no limit | 1 | ||||
2 | 1.5 A/20 V | DC,no limit | 1 | ||||
3 | 3 A/6 V | DC,no limit | 1 | ||||
4 | 3 A/20 V | 1ms | 0.1 | ||||
5 | 10 A/6 V | 1ms | 0.1 |
Typical output settling time
Source |
Range |
Output settling time |
Condition |
||
Fast 1,2 |
Normal 1 |
Slow 1 |
|||
Voltage |
200 V |
<1.3 ms |
<1.5 ms |
<2.5 ms |
Time required to reach within 0.1 % of final value at open load condition. Step is 10 % to 90 % range |
20 V |
<300 μs |
<360 μs |
<1 ms |
||
6 V |
<150 μs |
<250 μs |
<1 ms |
||
200 mV |
<200 μs |
<250 μs |
<1 ms |
||
Current |
3 A |
<200 μs |
<280 μs |
<1.2 ms |
Time required to reach within 0.1 % (0.3 % for 3 A range) of final value at short condition. Step is 10 % to 90 % range, Pulse current source base 6V voltage range and 105% range limit |
1.5 A |
<200 μs |
<280 μs |
<1.2 ms |
||
150 mA |
<200 μs |
<280 μs |
<1.2 ms |
||
15 mA |
<200 μs |
<280 μs |
<1.2 ms |
||
1.5 mA |
<200 μs |
<280 μs |
<1.2 ms |
||
150 μA |
<250 μs |
<300 μs |
<1.2 ms |
||
15 μA |
<250 μs |
<1.2 ms |
<2 ms |
||
1.5 μA |
<600 μs |
<1.2 ms |
<5 ms |
||
150 nA |
<600 μs |
<5 ms |
<12 ms |
I-V Out capability
Typical output settling time
Source |
Range |
Output settling time |
Condition |
||||
Fast1,2 | Normal1 | Slow1 | |||||
Voltage |
200V | <1.3ms | <1.5ms | <2.5ms |
Time required to reach within 0.1 % of final value at open load condition. Step is 10 % to 90 % range |
||
20V | <300μs | <360μs | <1ms | ||||
6V | <150μs | <250μs | <1ms | ||||
200mV | <200μs | <250μs | <1ms | ||||
Current |
3A | <200μs | <280μS | <1.2ms |
Time required to reach within 0.1 % (0.3 % for 3 A range) of final value at short condition. Step is 10 % to 90 % rangePulse current source base 6V voltage range and 105% range limit |
||
1.5A | <200μs | <280μS | <1.2ms | ||||
150mA | <200μs | <280μS | <1.2ms | ||||
15mA | <200μs | <280μs | <1.2ms | ||||
1.5mA | <200μs | <280μs | <1.2ms | ||||
150μA | <250μs | <300μs | <1.2ms | ||||
15μA | <250μs | <1.2ms | <2ms | ||||
1.5μA | <600μs | <1.2ms | <5ms | ||||
150nA | <600μs | <5ms | <12ms |
Sampling rate and NPLC setting
Setting | Range | ||||||
NPLC | 0.00005 PLC ~ 10 PLC | ||||||
Sampling Rate | 5 sps ~ 1 Msps |